Engineering Report Appendix 3.

Deviation of Stimulus Output Current from Device Intensity Settings

2000 Hz. Frequency Neurometer® CPT/C Medi-Dx 7000
1.5mA Knob Setting ­ 5,000 Ohm Load + 0.7% - 69.3%
1.5mA Knob Setting ­ 10,000 Ohm Load - 0.7% - 77.3%
9.99mA Knob Setting ­ 5,000 Ohm Load + 0.1% - 50.5%
9.99mA Knob Setting ­ 10,000 Ohm Load + 3.0% - 66.0%

250 Hz. Frequency Neurometer® CPT/C Medi-Dx 7000
1.5mA Knob Setting ­ 5,000 Ohm Load + 0.7% - 59.3%
1.5mA Knob Setting ­ 10,000 Ohm Load - 2.0% - 72.6%
9.99mA Knob Setting ­ 5,000 Ohm Load - 1.3% - 44.4%
9.99mA Knob Setting ­ 10,000 Ohm Load -2.0% - 62.0%

5 Hz. Frequency Neurometer® CPT/C Medi-Dx 7000
1.5mA Knob Setting ­ 5,000 Ohm Load - 2.7% - 42.0% / - 61.4% Asymmetric Output
1.5mA Knob Setting ­ 10,000 Ohm Load - 4.0% - 56.0% / - 74.0% Asymmetric Output
9.99mA Knob Setting ­ 5,000 Ohm Load - 4.3% - 31.3% / - 48.1% Asymmetric Output
9.99mA Knob Setting ­ 10,000 Ohm Load - 5.2% - 50.4% / - 66.4% Asymmetric Output



top
Use your browser's "Back" button to return to the previous page
or click on a link below to continue.

 /  Medi-Dx Truth?  /  Engineer Report  /  Procedures  / 

 /  News  /  Search  /  FAQ  /  Products  / 
 /  Abstracts & Bibliographies  /  Clinical Applications  / 
 /  Distributors  /  Contact Info  /  Documents & Downloads  /  Home  / 
 /  Legal & Privacy Notices  / 

Modified 10/05/00.
The entire contents of this website are ©Copyright 1996-2005. Neurotron, Incorporated, Baltimore, MD, USA. All rights reserved, worldwide.