Device Stimulus Output Comparison - Test 1
| Frequency Setting: | 2,000 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes, 2,000 Hertz and
the outputs are connected to a 5,000 Ohm resistor. The output voltage is measured
with the sampling rate set to 50,000 kHz, 50 of the 5,000 samples acquired are shown
here. The output current is obtained by dividing the output voltage by 5,000 Ohms. The
Neurometer® CPT/C output is a sine wave with a peak value of 1.5039 milliAmperes.
The Medi-Dx 7000 output is an undefined waveform with a peak value of 0.441 milliAmperes.
|
Device Stimulus Output Comparison - Test 2
| Frequency Setting: | 2,000 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 2,000
Hertz and the outputs are connected to a 5,000 Ohm resistor. The output
voltage is measured with the sampling rate set to 50,000 kHz, 50 of
the 5,000 samples acquired are shown here. The output voltage is obtained
by multiplying the measured voltage by the attenuation factor. Then
dividing the calculated output voltage by 5,000 Ohms to obtain the output
current (appendix #2). The Neurometer® CPT/C output is a sine wave
with a peak value of 9.96 milliAmperes. The Medi-Dx 7000 output is an
undefined waveform with a peak value of 4.47 milliAmperes.
|
Device Stimulus Output Comparison - Test 3
| Frequency Setting: | 250 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes, 250
Hertz and the outputs are connected to a 5,000 Ohm resistor. The output
voltage is measured with the sampling rate set to 25,000 kHz, 201 of
the 1200 samples acquired are shown here. The output current is obtained
by dividing the output voltage by 5,000 Ohms. The Neurometer® CPT/C
output is a sine wave with a peak value of 1.51 milliAmperes. The
Medi-Dx 7000 output is a square wave with a peak value of 0.605
milliAmperes.
|
Device Stimulus Output Comparison - Test 4
| Frequency Setting: | 250 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 250
Hertz and the outputs are connected to a 5,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 25,000 kHz,
201 of the 5,000 samples acquired are shown here. The output voltage
is obtained by multiplying the measured voltage by the attenuation
factor. Then dividing the calculated output voltage by 5,000 Ohms to
obtain the output current (appendix #2). The Neurometer® CPT/C
output is a sine wave with a peak value of 9.86 milliAmperes. The
Medi-Dx 7000 output is square wave with a peak value of 5.55 milliAmperes.
|
Device Stimulus Output Comparison - Test 5
| Frequency Setting: | 5 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes,
5 Hertz and the outputs are connected to a 5,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 8,000 kHz,
3,200 of the 5,000 samples acquired are shown here. The output
current is obtained by dividing the output voltage by 5,000 Ohms.
The Neurometer® CPT/C output is a sine wave with a peak value
of 1.46 milliAmperes. The Medi-Dx 7000 output is an undefined
waveform with a peak value of 0.874/0.580 milliAmperes (i.e. the
output is not symmetrical).
|
Device Stimulus Output Comparison - Test 6
| Frequency Setting: | 5 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 5,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 5
Hertz and the outputs are connected to a 5,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 8,000 kHz,
3,200 of the 5,000 samples acquired are shown here. The output
voltage is obtained by multiplying the measured voltage by the
attenuation factor. Then dividing the calculated output voltage by
5,000 Ohms to obtain the output current (appendix #2). The
Neurometer® CPT/C output is a sine wave with a peak value of
9.56 milliAmperes. The Medi-Dx 7000 output is an undefined waveform
with a peak value of 6.86/5.17 milliAmperes (i.e. the output is not
symmetrical).
|
Device Stimulus Output Comparison - Test 7
| Frequency Setting: | 2000 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes, 2,000
Hertz and the outputs are connected to a 10,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 50,000 kHz,
50 of the 5,000 samples acquired are shown here. The output voltage is
obtained by multiplying the measured voltage by the attenuation
factor. Then dividing the calculated output voltage by 10,000 Ohms to
obtain the output current (appendix #2). The Neurometer® CPT/C
output is a sine wave with a peak value of 1.49 milliAmperes. The
Medi-Dx 7000 output is an undefined waveform with a peak value of
0.335 milliAmperes.
|
Device Stimulus Output Comparison - Test 8
| Frequency Setting: | 2000 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 2,000
Hertz and the outputs are connected to a 10,000 Ohm resistor. The output
voltage is measured with the sampling rate set to 50,000 kHz, 50 of the
5,000 samples acquired are shown here. The output voltage is obtained by
multiplying the measured voltage by the attenuation factor. Then dividing
the calculated output voltage by 10,000 Ohms to obtain the output current
(appendix #2). The Neurometer® CPT/C output is a sine wave with a
peak value of 10.29 milliAmperes. The Medi-Dx 7000 output is an undefined
waveform with a peak value of 3.39 milliAmperes.
|
Device Stimulus Output Comparison - Test 9
| Frequency Setting: | 250 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes, 250
Hertz and the outputs are connected to a 10,000 Ohm resistor. The output
voltage is measured with the sampling rate set to 25,000 kHz, 201 of
the 1,200 samples acquired are shown here. The output voltage is obtained
by multiplying the measured voltage by the attenuation factor. Then
dividing the calculated output voltage by 10,000 Ohms to obtain the output
current (appendix #2). The Neurometer® CPT/C output is a sine wave
with a peak value of 1.47 milliAmperes. The Medi-Dx 7000 output is a
square wave with a peak value of 0.410 milliAmperes.
|
Device Stimulus Output Comparison - Test 10
| Frequency Setting: | 250 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 250
Hertz and the outputs are connected to a 10,000 Ohm resistor. The output
voltage is measured with the sampling rate set to 25,000 kHz, 201 of the
1,200 samples acquired are shown here. The output voltage is obtained by
multiplying the measured voltage by the attenuation factor. Then dividing
the calculated output voltage by 10,000 Ohms to obtain the output current
(appendix #2). The Neurometer® CPT/C output is a sine wave with a peak
value of 9.79 milliAmperes. The Medi-Dx 7000 output is a square wave
with a peak value of 3.80 milliAmperes.
|
Device Stimulus Output Comparison - Test 11
| Frequency Setting: | 5 Hz. |
| Intensity Setting: | 1.50 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 1.5 milliAmperes, 5
Hertz and the outputs are connected to a 10,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 8,000 kHz,
3,200 of the 5,000 samples acquired are shown here. The output voltage
is obtained by multiplying the measured voltage by the attenuation
factor. Then dividing the calculated output voltage by 10,000 Ohms
to obtain the output current (appendix #2). The Neurometer®
CPT/C output is a sine wave with a peak value of 1.44 milliAmperes.
The Medi-Dx 7000 output is an undefined waveform with a peak value
of 0.662/0.385 milliAmperes (i.e. the output is not symmetrical).
|
Device Stimulus Output Comparison - Test 12
| Frequency Setting: | 5 Hz. |
| Intensity Setting: | 9.99 mAmps. |
| Skin Simulation Load: | 10,000 Ohms |
| Neurometer® CPT/C Stimulus |
Medi-Dx 7000 Stimulus |
 |
 |
|
The output controls for each unit are set for 9.99 milliAmperes, 5
Hertz and the outputs are connected to a 10,000 Ohm resistor. The
output voltage is measured with the sampling rate set to 8,000 kHz,
3,200 of the 5,000 samples acquired are shown here. The output voltage
is obtained by multiplying the measured voltage by the attenuation
factor. Then dividing the calculated output voltage by 10,000 Ohms to
obtain the output current (appendix #2). The Neurometer® CPT/C
output is a sine wave with a peak value of 9.47 milliAmperes. The
Medi-Dx 7000 output is an undefined waveform with a peak value of
4.96/3.36 milliAmperes (i.e. the output is not symmetrical).
|
[top]